An Improved Design and Simplified Evaluation Technique for Waveguide Microcalorimeter

An improved design and simplified evaluation technique for waveguide microcalorimeter is presented in this paper. The design is based on a traditional twin-line microcalorimeter for calibrating thermistor mounts but with a proposed novel modification. In the improved design, another thermistor is ad...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 2016-06, Vol.65 (6), p.1450-1455
Hauptverfasser: Cui, Xiaohai, Meng, Yu Song, Li, Yong, Zhang, Yue, Shan, Yueyan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An improved design and simplified evaluation technique for waveguide microcalorimeter is presented in this paper. The design is based on a traditional twin-line microcalorimeter for calibrating thermistor mounts but with a proposed novel modification. In the improved design, another thermistor is added to measure the temperature change of the thermal isolation waveguide section. A simplified evaluation method is also proposed to minimize the S-parameter measurements for obtaining the effective efficiency of the mount. The measurement model and its detailed derivation are discussed. The capability of the improved design and proposed evaluation technique has been assessed through a recent consultative committee for electricity and magnetism key comparison.
ISSN:0018-9456
1557-9662
DOI:10.1109/TIM.2016.2534239