Calibration of the PTP vector network analyzer

A calibration procedure for perturbation two-port vector network analyzer (PTP VNA) is presented. PTP VNA consists of a variable perturbation two-port (PTP) placed between a device under test (DUT) and a scalar network analyzer (SNA). A vector reflection coefficient of the DUT is determined on the b...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Hoffmann, K., Skvor, Z.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A calibration procedure for perturbation two-port vector network analyzer (PTP VNA) is presented. PTP VNA consists of a variable perturbation two-port (PTP) placed between a device under test (DUT) and a scalar network analyzer (SNA). A vector reflection coefficient of the DUT is determined on the basis of amplitude only measurements. A full correction of SNA systematic errors is possible. In principle it may be used up to millimetre and submillimetre frequency bands.
DOI:10.1109/MIKON.1998.742810