Traceability improvement of high temperature thermal property measurements of materials for new fission reactors

New metrological facilities for high temperature measurements of thermophysical properties have been recently developed by European National Metrology Institutes (NMIs). This work, performed in the framework of the joint research project "MetroFission", aims to provide reliable measurement...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Hay, B., Anhalt, K., Chapman, L., Boboridis, K., Hameury, J., Krenek, S., Vlahovic, L., Fleurence, N., Benes, O.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:New metrological facilities for high temperature measurements of thermophysical properties have been recently developed by European National Metrology Institutes (NMIs). This work, performed in the framework of the joint research project "MetroFission", aims to provide reliable measurements up to 2000 °C for materials having thermal properties similar to those required for design of new generations of fission reactors. This will enable to improve the traceability to the International System of Units (SI) of measurements performed by nuclear research laboratories. The validation of these reference apparatuses was done at high temperature by measuring the thermal diffusivity and normal spectral emissivity of an isostatic graphite and pure tungsten.
DOI:10.1109/ANIMMA.2013.6727928