Avalanche detector with ultraclean response for time-resolved photon counting
Experimental tests have been carried out for characterizing the performance of a new single-photon avalanche diode. The detector is specifically designed for picosecond time-correlated single photon counting, aiming to obtain a time response free from the tail effects and/or secondary bumps observed...
Gespeichert in:
Veröffentlicht in: | IEEE journal of quantum electronics 1998-05, Vol.34 (5), p.817-821 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Experimental tests have been carried out for characterizing the performance of a new single-photon avalanche diode. The detector is specifically designed for picosecond time-correlated single photon counting, aiming to obtain a time response free from the tail effects and/or secondary bumps observed in all other available single-photon detectors. The experimental results confirm that an unprecedented ultraclean response is attained, with full width at half maximum (FWHM) of 35 ps, full width at a thousandth of the peak count of 214 ps, together with a photon detection efficiency of 35% around 580-nm wavelength. |
---|---|
ISSN: | 0018-9197 1558-1713 |
DOI: | 10.1109/3.668769 |