Passive Shimming for Magic-Angle-Spinning NMR
The superconducting dipole magnets wound with HTS wires have been developed for a magic-angle-spinning NMR. The magnetic field of the dipole magnets is tilted from z-axis. Since the highly homogeneous magnetic field is required, the magnetic field of the dipole magnets is necessarily compensated by...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on applied superconductivity 2014-06, Vol.24 (3), p.1-4 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The superconducting dipole magnets wound with HTS wires have been developed for a magic-angle-spinning NMR. The magnetic field of the dipole magnets is tilted from z-axis. Since the highly homogeneous magnetic field is required, the magnetic field of the dipole magnets is necessarily compensated by passive and/or active shimming. However, the compensation of all the components has not been reported. Usually, due to the axial symmetry of a solenoid magnet, only z component of the magnetic field is homogenized. In this paper, all the x, y, and z components of the magnetic field generated by the superconducitng dipole magnet are compensated because of the tilt from the z-axis. |
---|---|