Ultrasound properties and images of ex-vivo liver during thermal therapy

The authors have measured changes in ex-vivo bovine liver ultrasound backscatter coefficient and attenuation coefficient as a function of time at various temperatures below the vaporization threshold of water (100/spl deg/C), to predict the appearance of real-time ultrasound images of tissues during...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Gertner, M.R., Wilson, B.C., Worthington, A.E., Sherar, M.D.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The authors have measured changes in ex-vivo bovine liver ultrasound backscatter coefficient and attenuation coefficient as a function of time at various temperatures below the vaporization threshold of water (100/spl deg/C), to predict the appearance of real-time ultrasound images of tissues during thermal therapy. The authors have also used a clinical ultrasound scanner to record B-scan images of ex-vivo bovine liver heated with an interstitial microwave antenna to mimic a clinical treatment. Image echogenicity decreased with heating, likely due to increased attenuation in the coagulated tissue.
ISSN:1051-0117
DOI:10.1109/ULTSYM.1997.661827