Quantitative performance assessments for neuromagnetic imaging systems

We have developed a Monte-Carlo simulation method to assess the performance of neuromagnetic imaging systems using two kinds of performance metrics: A-prime metric and spatial resolution. We compute these performance metrics for virtual sensor systems having 80, 160, 320, and 640 sensors, and discus...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Koga, Ryo, Hiyama, Ei, Matsumoto, Takuya, Sekihara, Kensuke
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We have developed a Monte-Carlo simulation method to assess the performance of neuromagnetic imaging systems using two kinds of performance metrics: A-prime metric and spatial resolution. We compute these performance metrics for virtual sensor systems having 80, 160, 320, and 640 sensors, and discuss how the system performance is improved, depending on the number of sensors. We also compute these metrics for existing whole-head MEG systems, MEGvision™ (Yokogawa Electric Corporation, Tokyo, Japan) that uses axial-gradiometer sensors, and TRIUX™ (Elekta Corporate, Stockholm, Sweden) that uses planar-gradiometer and magnetometer sensors. We discuss performance comparisons between these significantly different systems.
ISSN:1094-687X
1557-170X
1558-4615
DOI:10.1109/EMBC.2013.6610524