Effect of impeller flow path on pump performance and impeller stability of the monopivot circulatory pump

The effect of a cutout on the pump pressure-flow characteristics and the impeller stability was quantified using computational fluid dynamics analysis in order to provide good hemocompatibility of the monopivot extracorporeal circulation pump. As a result, the following findings were clarified. The...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) 2013-01, Vol.2013, p.2736-2739
Hauptverfasser: Nishida, Masahiro, Nakayama, Kento, Kosaka, Ryo, Maruyama, Osamu, Kawaguchi, Yasuo, Kuwana, Katsuyuki, Yamane, Takashi
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The effect of a cutout on the pump pressure-flow characteristics and the impeller stability was quantified using computational fluid dynamics analysis in order to provide good hemocompatibility of the monopivot extracorporeal circulation pump. As a result, the following findings were clarified. The pump pressure is lower in the cutout model than in the no-cutout model. The impeller stability with respect to the buoyancy of the impeller is better in the cutout model than in the no-cutout model. The impeller stability with respect to the impeller tilt is better in the cutout model than in the no-cutout model. Therefore, the cutout model, in which the geometry corresponds to the commercialized pump, was likely to be better than the no-cutout model because the stability that has the possibility to decrease the gap instantaneously to increase hemolysis despite the impeller rotational speed slightly.
ISSN:1094-687X
1557-170X
1558-4615
DOI:10.1109/EMBC.2013.6610106