NPLI Cesium Atomic Fountain Frequency Standard: Preliminary Results

National Physical Laboratory India has developed a cesium atomic fountain frequency standard. The fountain has now become operational and is ready for complete frequency evaluation. In this paper, we report for the first time, the preliminary results obtained with this fountain, which include Ramsey...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 2013-07, Vol.62 (7), p.2037-2042
Hauptverfasser: Arora, P., Purnapatra, S. B., Acharya, A., Agarwal, A., Yadav, S., Pant, K., Gupta, A. S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:National Physical Laboratory India has developed a cesium atomic fountain frequency standard. The fountain has now become operational and is ready for complete frequency evaluation. In this paper, we report for the first time, the preliminary results obtained with this fountain, which include Ramsey fringes, frequency stability analysis, with an H-maser, and magnetic field mapping in the atomic flight region.
ISSN:0018-9456
1557-9662
DOI:10.1109/TIM.2013.2255974