Design of Wavelet-Based ECG Detector for Implantable Cardiac Pacemakers

A wavelet Electrocardiogram (ECG) detector for low-power implantable cardiac pacemakers is presented in this paper. The proposed wavelet-based ECG detector consists of a wavelet decomposer with wavelet filter banks, a QRS complex detector of hypothesis testing with wavelet-demodulated ECG signals, a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on biomedical circuits and systems 2013-08, Vol.7 (4), p.426-436
Hauptverfasser: Min, Young-Jae, Kim, Hoon-Ki, Kang, Yu-Ri, Kim, Gil-Su, Park, Jongsun, Kim, Soo-Won
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A wavelet Electrocardiogram (ECG) detector for low-power implantable cardiac pacemakers is presented in this paper. The proposed wavelet-based ECG detector consists of a wavelet decomposer with wavelet filter banks, a QRS complex detector of hypothesis testing with wavelet-demodulated ECG signals, and a noise detector with zero-crossing points. In order to achieve high detection accuracy with low power consumption, a multi-scaled product algorithm and soft-threshold algorithm are efficiently exploited in our ECG detector implementation. Our algorithmic and architectural level approaches have been implemented and fabricated in a standard 0.35 μm CMOS technology. The testchip including a low-power analog-to-digital converter (ADC) shows a low detection error-rate of 0.196% and low power consumption of 19.02 μW with a 3 V supply voltage.
ISSN:1932-4545
1940-9990
DOI:10.1109/TBCAS.2012.2229463