The novel research of intraocular pressure tonometer by using inductance sensor

This article mainly discusses the theoretical derivation and the estimate of the intraocular pressure (IOP) sensor by using inductance. The inductance changes associate with the variation of anterior corneal curvature radius due to the differences of IOP. The value of inductance is transformed by th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Yu-Shun Tang, Wei-De Jeng, Ting-Wei Huang, Mang Ou-Yang, Jin-Chern Chiou, Jeng-Ren Duann, Ching-Hsing Luo, Hong-Yi Huang, Yi-Wu Tsai
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This article mainly discusses the theoretical derivation and the estimate of the intraocular pressure (IOP) sensor by using inductance. The inductance changes associate with the variation of anterior corneal curvature radius due to the differences of IOP. The value of inductance is transformed by the LC oscillator, and the signal is transmitted by antenna, and finally read by the exterior readout circuit to achieve IOP measuring. The result show that the estimative values of inductance is close to the measured values, and the percentage of inductance variation is about 0.08% per 1 mmHg. Because of the low percentage, it makes much harder to realize readout circuit. In the future, the alteration of structure and material are considered to increase the percentage of varying inductance.
ISSN:1930-0395
2168-9229
DOI:10.1109/ICSENS.2012.6411158