TlBr Capacitive Frisch Grid Detectors
Capacitive Frisch grid detectors with the dimensions of 2 mm × 2 mm × 4.4 mm were fabricated from TlBr crystals grown by the travelling molten zone method using zone-purified materials. Spectroscopic performance of the detectors was evaluated at room temperature (24°C), - 5°C and - 20°C. The TlBr de...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on nuclear science 2013-04, Vol.60 (2), p.1156-1161 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Capacitive Frisch grid detectors with the dimensions of 2 mm × 2 mm × 4.4 mm were fabricated from TlBr crystals grown by the travelling molten zone method using zone-purified materials. Spectroscopic performance of the detectors was evaluated at room temperature (24°C), - 5°C and - 20°C. The TlBr detector exhibited an energy resolution of 1.6% FWHM at 662 keV with a peak-to-Compton ratio of 8.7 with the depth correction at room temperature. Improvement of the detector performance was achieved by cooling the detector-preamplifier system. The energy resolutions of 3.4%, 1.5%, 1.0% and 0.9% FWHM at 122 keV, 356 keV, 662 keV and 1332 keV, respectively, were recorded with the TlBr detector without the depth correction at - 5°C. An energy resolution of 0.9% FWHM at 662 keV with a peak-to-Compton ratio of 18.2 was obtained from the TlBr detector with the depth correction at -20°C. An upper limit for the Fano factor of TlBr was estimated to be 0.43 at -20°C. |
---|---|