An automated set-up for calibration of inductive voltage dividers
An automated system has been built for calibration of inductive voltage dividers (IVDs) in the frequency range from 50 Hz to 10 kHz. In this system, computer controlled IVDs are used to automatically find the in-phase and quadrature bridge balance. Initial tests at 1 kHz have shown an agreement with...
Gespeichert in:
Hauptverfasser: | , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An automated system has been built for calibration of inductive voltage dividers (IVDs) in the frequency range from 50 Hz to 10 kHz. In this system, computer controlled IVDs are used to automatically find the in-phase and quadrature bridge balance. Initial tests at 1 kHz have shown an agreement with results from previous calibrations within the estimated uncertainties of 1.0 × 10 -7 V/V in-phase and 1.0 × 10 -6 V/V quadrature. |
---|---|
ISSN: | 0589-1485 2160-0171 |