Bilateral comparison on calibration factor of a 2.92 mm power sensor
This paper describes a bilateral comparison between PTB and SP on the calibration factor of a 2.92 mm thermocouple power sensor. The measurements were done at 0 dBm from 50 MHz to 40 GHz. The comparison also covers the sensor reflection coefficient over the same frequency range.