Spatial compounding of trans-esophageal echo volumes using X-ray probe tracking

The use of ultrasound imaging for image guidance of cardiac procedures is limited by the small field of view of the ultrasound volume. A larger view can be created by image-based registration of partially overlapping volumes, but automatic registration often fails unless the volume alignment is init...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Housden, R. J., Ma, Y., Arujuna, A., Nijhof, N., Cathier, P., Gijsbers, G., Bullens, R., Rhode, K. S.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The use of ultrasound imaging for image guidance of cardiac procedures is limited by the small field of view of the ultrasound volume. A larger view can be created by image-based registration of partially overlapping volumes, but automatic registration often fails unless the volume alignment is initialised close to the volumes' correct alignment. In this paper, we use X-ray images to track a trans-esophageal probe, which provides the required position estimates for the ultrasound volumes. The tracking is possible using multiple X-rays or just one X-ray at each probe position. We test the method in a phantom experiment and find that with at least 50% volume overlap, 98% of ultrasound volume pairs are correctly registered when tracked using three X-rays per volume, and 95% using single X-rays.