Electrical performance of penetrating microelectrodes chronically implanted in cat cortex

Penetrating multielectrode arrays with electrode coatings of sputtered iridium oxide (SIROF) have been implanted chronically in cat cortex for periods over 300 days. The ability of these electrodes to inject charge at levels above expected thresholds for neural excitation has been examined in vivo b...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2011-01, Vol.2011, p.5416-5419
Hauptverfasser: Kane, S. R., Cogan, S. F., Ehrlich, J., Plante, T. D., McCreery, D. B.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Penetrating multielectrode arrays with electrode coatings of sputtered iridium oxide (SIROF) have been implanted chronically in cat cortex for periods over 300 days. The ability of these electrodes to inject charge at levels above expected thresholds for neural excitation has been examined in vivo by measurements of voltage transients in response to current-controlled, cathodal stimulation pulsing. The effect of current pulse width from 150 (is to 500 (is and voltage biasing of the electrodes in the interpulse period at two levels, 0.0 V and 0.6 V vs. Ag|AgCl, were also investigated. The results of in vivo characterization of the electrodes by open-circuit potential measurements, cyclic voltammetry and impedance spectroscopy are also reported.
ISSN:1094-687X
1557-170X
1558-4615
DOI:10.1109/IEMBS.2011.6091339