Structure From Flow for Calibration of an Array Air-Flow Sensor
We describe the automatic calibration of a set of air-flow sensitive sensors exposed to a set of unknown and randomly distributed air-flow stimuli. The algorithm makes use of the singular value decomposition of the sensor data and the known sensor model. The sensor array configuration consists of th...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 2012-02, Vol.61 (2), p.307-315 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe the automatic calibration of a set of air-flow sensitive sensors exposed to a set of unknown and randomly distributed air-flow stimuli. The algorithm makes use of the singular value decomposition of the sensor data and the known sensor model. The sensor array configuration consists of the maximum response direction vectors of the individual air-flow sensors. The relative orientations of the direction vectors of the array elements are automatically determined. The orientation of the entire array in a world reference frame has to be manually calibrated. The change in the air flow can be measured without any manual calibration. |
---|---|
ISSN: | 0018-9456 1557-9662 |
DOI: | 10.1109/TIM.2011.2164286 |