An investigation of alternating impedance microstrip transceiver coil arrays for MRI at 7T

The magnetic field generated from the conventional microstrip resonator (MSR) is inhomogeneous along the resonator axis. It is strong at the centre and weak at the terminating ends. To homogenize the B 1 + field, it is required to modify the traditional MSR. Recently, an approach based on alternatin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Elabyad, I. A., Omar, A.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The magnetic field generated from the conventional microstrip resonator (MSR) is inhomogeneous along the resonator axis. It is strong at the centre and weak at the terminating ends. To homogenize the B 1 + field, it is required to modify the traditional MSR. Recently, an approach based on alternating impedance MSR has been presented. From this contribution, another configuration of alternating impedance resonator with circular elements is proposed. Electromagnetic (EM) simulations were performed for three types of MSR volume coils (one traditional and two alternating impedance) for MRI at 7T. Results indicate that, the alternating impedance MSR with circular elements is the best coil in terms of the B 1 + field homogeneity and the E/B 1 + field ratio reduction at 7T.
ISSN:0149-645X
2576-7216
DOI:10.1109/MWSYM.2011.5972857