Cardioversion of atrial fibrillation (AF) can be achieved by applying an electric shock to the heart (defibrillation). A cost-effective approach for internal cardioversion is by means of a subcutaneous passive implantable atrial defibrillator (PIAD). A circuit design using a dual tuned resonant circ...
Gespeichert in:
Hauptverfasser: | , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cardioversion of atrial fibrillation (AF) can be achieved by applying an electric shock to the heart (defibrillation). A cost-effective approach for internal cardioversion is by means of a subcutaneous passive implantable atrial defibrillator (PIAD). A circuit design using a dual tuned resonant circuit whose capacitance on the receiver side also serves as a voltage booster is proposed and evaluated. The circuit performance is characterised using a MATLAB based model and a PSpice simulation. The circuit exhibited inherent voltage gain stability with a reasonable voltage gain (V Rx /V Tx >; 0.3). Computed link efficiency modelling, estimated high link efficiency values of about 63.63% at the 208 kHz resonant frequency (26 mm air gap between coils). The above theoretical characterisation was supported by hardware prototyping which realised a system efficiency of 59.61% (best case). |
---|---|