Establishment of 40,000 A Rogowski coil calibration system and its uncertainty analysis
System for calibrating Rogowski coil(RC) up to primary current of 40,000 A have been established by employing standard current transformer(CT), CT comparator and voltage to current converter(VCC). The expanded uncertainty is 0.021 % ~ 0.230 % for ratio error and 0.3 min ~ 1.9 min for phase displacem...
Gespeichert in:
Hauptverfasser: | , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | System for calibrating Rogowski coil(RC) up to primary current of 40,000 A have been established by employing standard current transformer(CT), CT comparator and voltage to current converter(VCC). The expanded uncertainty is 0.021 % ~ 0.230 % for ratio error and 0.3 min ~ 1.9 min for phase displacement in the primary current ranges of 10 ~ 40,000 A. |
---|---|