A heterodyne receiver for harmonic doppler radar cardio-pulmonary monitoring with body-worn passive RF tags

Harmonic Doppler radar can be used to isolate a subject's respiratory motion from environmental clutter. In this paper, a heterodyne receiver is used to detect tag motion and human cardiopulmonary activity, with the results compared with a that using a homodyne receiver. The heterodyne system i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Singh, A., Lubecke, V. M.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Harmonic Doppler radar can be used to isolate a subject's respiratory motion from environmental clutter. In this paper, a heterodyne receiver is used to detect tag motion and human cardiopulmonary activity, with the results compared with a that using a homodyne receiver. The heterodyne system is shown to work at a greater distance from the subject (more than 160 cm) with greater sensitivity to tag motion (more than 10 dB), using the same transmitted power (10 dBm). Both homodyne and heterodyne receivers are also used to successfully detect a human heartbeat by attaching a passive harmonic tag over a subject's chest.
ISSN:0149-645X
2576-7216
DOI:10.1109/MWSYM.2010.5515845