Biopsy needle tract cauterization using an embedded array of piezoceramic microheaters
This paper presents the use of an array of piezoceramic microheaters embedded in a biopsy needle for tract cauterization. Circular PZT-5A heaters of 200 ¿m diameter and 70-80 ¿m thickness are embedded in the wall of a 20-gauge stainless steel needle. Finite element modeling suggests that a PZT array...
Gespeichert in:
Hauptverfasser: | , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This paper presents the use of an array of piezoceramic microheaters embedded in a biopsy needle for tract cauterization. Circular PZT-5A heaters of 200 ¿m diameter and 70-80 ¿m thickness are embedded in the wall of a 20-gauge stainless steel needle. Finite element modeling suggests that a PZT array with no gap between the elements to be the most suitable design. The temperature profile generated is measured at two resonance modes: the radial mode (10.3 MHz) and the thickness mode (22.3 MHz). The needle surface exceeds the minimum target temperature rise of 33°C for applied voltages of 17 V RMS and 14 VRMS for radial and thickness mode, respectively. The corresponding input powers are 236 mW and 325 mW, respectively. The tissue cauterization extends 1-1.25 mm beyond the perimeter of needle. |
---|---|
ISSN: | 1084-6999 |
DOI: | 10.1109/MEMSYS.2010.5442374 |