The XAMPS detector for the X-ray Pump-Probe instrument at LCLS
An integrating fast readout detector is being developed for the X-ray Pump Probe (XPP) instrument at the Linac Coherent Light Source (LCLS). The detector is based on X-ray Active Matrix Pixel Sensors (XAMPS) with their dedicated readout ASICs. It covers a dynamic range of more than 10,000 photons of...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An integrating fast readout detector is being developed for the X-ray Pump Probe (XPP) instrument at the Linac Coherent Light Source (LCLS). The detector is based on X-ray Active Matrix Pixel Sensors (XAMPS) with their dedicated readout ASICs. It covers a dynamic range of more than 10,000 photons of 8 keV energy with a resolution better than half photon FWHM. The frame of 1024 × 1024 pixels will be read out within the repetition rate of the X-ray laser pulses which will be as fast as 120 Hz. The structure of the sensor allows full fill factor and -100% efficiency at the energy of interest. |
---|---|
ISSN: | 1082-3654 2577-0829 |