Measurement of the MTF of a Cone-Beam Breast Computed Tomography laboratory scanner
In the framework of the realization of dedicated Cone-Beam Breast Computed Tomography (CBBCT) apparatus we report the MTF characterization of our prototype having the best performance in terms of detector pixel size and X ray focal spot than any other systems built or under construction. From our me...
Gespeichert in:
Hauptverfasser: | , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In the framework of the realization of dedicated Cone-Beam Breast Computed Tomography (CBBCT) apparatus we report the MTF characterization of our prototype having the best performance in terms of detector pixel size and X ray focal spot than any other systems built or under construction. From our measurement we estimated a value of the longitudinal and transverse resolution of 4-5 lp/mm (at 10% MTF). The presence of CT image features has been observed and attributed of specific phase contrast effects. |
---|---|