Imaging performance of PEM-1: a high resolution system for positron emission mammography
Positron emission tomography with /sup 18/F-fluoro-deoxyglucose (FDG) is known to detect the increased metabolism of breast tumours. The authors have built a positron emission mammography system which produces metabolic breast images co-registered with conventional mammography. Two position sensitiv...
Gespeichert in:
Hauptverfasser: | , , , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Positron emission tomography with /sup 18/F-fluoro-deoxyglucose (FDG) is known to detect the increased metabolism of breast tumours. The authors have built a positron emission mammography system which produces metabolic breast images co-registered with conventional mammography. Two position sensitive photomultipliers are coupled to four BGO blocks which are partially cut into 2/spl times/2 mm elements from both faces. This provides 72/spl times/72 BGO crystals in coincidence yielding over 26 million lines of response. A real time display is formed by performing a weighted back-projection onto seven image planes through the breast. The prototype instrument is currently being tested on phantoms. The spatial resolution in the central region has been measured at 2.05 mm FWHM, less than half that of a conventional PET scanner. The coincidence resolving time is 12 nsec. This device should be able to identify breast tumours |
---|---|