Flexible quadrature coil system for simultaneous reception of magnetic resonance signals [biomedical NMR]

The magnetic field intensity of a flexible quadrature coil system for magnetic resonance imaging is modeled and simulated. The calculated signal to noise ratio of the coil system is verified experimentally using a coil constructed with the model's dimensions. It is found that flexing the coil o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on magnetics 1995-11, Vol.31 (6), p.3590-3592
Hauptverfasser: Molyneaux, D.A., Qureshi, A.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The magnetic field intensity of a flexible quadrature coil system for magnetic resonance imaging is modeled and simulated. The calculated signal to noise ratio of the coil system is verified experimentally using a coil constructed with the model's dimensions. It is found that flexing the coil over a thirty degree range preserves the high signal to noise ratio and increases the clinical utility of such a system.
ISSN:0018-9464
1941-0069
DOI:10.1109/20.489579