Electromagnetic interference effect from power line noise in electrocardiograph signal using faraday cage

The main objective of this study is to build a basic human size faraday cage. 1-channel ECG circuit inclusive of notch filter also has been built to test the effect of faraday cage. The outcome of ECG signals was compared when placed inside cage (inside building), outside faraday cage (inside buildi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Zakaria, N.A., Sudirman, R., Jamaluddin, M.N.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The main objective of this study is to build a basic human size faraday cage. 1-channel ECG circuit inclusive of notch filter also has been built to test the effect of faraday cage. The outcome of ECG signals was compared when placed inside cage (inside building), outside faraday cage (inside building) and outside building. Then the study of relationship between faraday cage and the effect on power line noise in ECG system has been done. From the result of this project it can be seen that faraday cage can eliminate the 50 Hz power line noise in ECG signal.
DOI:10.1109/PECON.2008.4762559