Electromagnetic interference to infusion pumps. Update2008 from GSM mobile phones

Electromagnetic interference (EMI) to critical care medical devices has been reported by various groups. Previuos study demonstrated that infusion pumps are susceptible of false alarm buzzing and block of infusion, when exposed to various EMI sources. Aim of this paper is to investigate the changes...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:2008 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2008-01, Vol.2008, p.4503-4506
Hauptverfasser: Calcagnini, G., Censi, F., Triventi, M., Mattei, E., LoSterzo, R., Marchetta, E., Bartolini, P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Electromagnetic interference (EMI) to critical care medical devices has been reported by various groups. Previuos study demonstrated that infusion pumps are susceptible of false alarm buzzing and block of infusion, when exposed to various EMI sources. Aim of this paper is to investigate the changes in the risk of EMI from the estimates of our previous 2005 survey and to extend the EMI risk assessment to newer telecommunication products: DECT phones and WiFi terminals. With regards to GSM phones, compare to the results obtained in 2005, we observed a decrease in the rate of failure (from 58% to 30%). From our findings, the use of WiFi and DECT does not pose a real risk to infusion systems.
ISSN:1094-687X
1557-170X
1558-4615
DOI:10.1109/IEMBS.2008.4650213