Proton beam energy measurement using semiconductor detectors at the 45MeV test beam line of PEFP
For the pilot and feasibility studies on the development of beam utilization technologies of PEFP (proton engineering frontier project), the test beam line was installed at the MC-50 cyclotron of KIRAMS (Korea institute of radiological and medical sciences) [1-2]. The energy measurement of proton be...
Gespeichert in:
Hauptverfasser: | , , , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | For the pilot and feasibility studies on the development of beam utilization technologies of PEFP (proton engineering frontier project), the test beam line was installed at the MC-50 cyclotron of KIRAMS (Korea institute of radiological and medical sciences) [1-2]. The energy measurement of proton beam with high accuracy is very important for the some experiments such as radiation hardness test of semiconductor devices for the spacecraft, detector development and test for the nuclear physics, etc. Energy measurement of 35 MeV and 45 MeV proton beam using a 5 mm thick Si(Li) in air was performed. The energy was controlled by Al degrader in the range of 0.02 mm~6.2 mm. The measured value was compared to the results of code simulation using SRIM 2003. |
---|---|
ISSN: | 1944-4680 2152-9582 |
DOI: | 10.1109/PAC.2007.4439969 |