SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus

Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Pisa, S., Calcagnini, G., Cavagnaro, M., Piuzzi, E., Triventi, M., Bernardi, P.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1757
container_issue
container_start_page 1754
container_title
container_volume
creator Pisa, S.
Calcagnini, G.
Cavagnaro, M.
Piuzzi, E.
Triventi, M.
Bernardi, P.
description Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm 3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6 degC for 6 minute MRI investigations. However, lower temperature increments are expected if blood per fusion and thermoregulation are considered
doi_str_mv 10.1109/MWSYM.2006.249384
format Conference Proceeding
fullrecord <record><control><sourceid>ieee_6IE</sourceid><recordid>TN_cdi_ieee_primary_4015289</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><ieee_id>4015289</ieee_id><sourcerecordid>4015289</sourcerecordid><originalsourceid>FETCH-LOGICAL-i175t-c827f88cf3cbf2bce2d3abba9c9e76889d0ed7e69cdb9bf0c10c29e872f82b773</originalsourceid><addsrcrecordid>eNotjN1KwzAYQAMqOOceQLzJC3R-SX-SXJbRucGKY5uoN478fIVqu5akE_f2ThQOHDgXh5A7BlPGQD2UL9u3csoBsilPVCyTCzJRQsKZWKUJY5dkBCxRUZakr9fkJoQPAEgly0bkfZtv6A7bHr0ejh5p0eCXHuruEGh9oGttsdWf6Omiaxz6QIvvvgvo6NDRoqTzGhsX6Np37mjP1ZxouVnSvO_17y_ckqtKNwEn_x6T53mxmy2i1dPjcpavopqJdIis5KKS0laxNRU3FrmLtTFaWYUik1I5QCcwU9YZZSqwDCxXKAWvJDdCxGNy__etEXHf-7rV_rRPgKVcqvgHp2ZVGA</addsrcrecordid><sourcetype>Publisher</sourcetype><iscdi>true</iscdi><recordtype>conference_proceeding</recordtype></control><display><type>conference_proceeding</type><title>SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus</title><source>IEEE Electronic Library (IEL) Conference Proceedings</source><creator>Pisa, S. ; Calcagnini, G. ; Cavagnaro, M. ; Piuzzi, E. ; Triventi, M. ; Bernardi, P.</creator><creatorcontrib>Pisa, S. ; Calcagnini, G. ; Cavagnaro, M. ; Piuzzi, E. ; Triventi, M. ; Bernardi, P.</creatorcontrib><description>Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm 3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6 degC for 6 minute MRI investigations. However, lower temperature increments are expected if blood per fusion and thermoregulation are considered</description><identifier>ISSN: 0149-645X</identifier><identifier>ISBN: 9780780395411</identifier><identifier>ISBN: 0780395417</identifier><identifier>DOI: 10.1109/MWSYM.2006.249384</identifier><language>eng</language><publisher>IEEE</publisher><subject>Cardiac pacemakers ; Catheters ; Head ; Humans ; Magnetic field measurement ; Magnetic resonance imaging ; Pacemakers ; Power distribution ; Pulse measurements ; Radio frequency ; Temperature ; Temperature distribution ; Thorax</subject><ispartof>2006 IEEE MTT-S International Microwave Symposium Digest, 2006, p.1754-1757</ispartof><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://ieeexplore.ieee.org/document/4015289$$EHTML$$P50$$Gieee$$H</linktohtml><link.rule.ids>309,310,778,782,787,788,2054,27908,54903</link.rule.ids><linktorsrc>$$Uhttps://ieeexplore.ieee.org/document/4015289$$EView_record_in_IEEE$$FView_record_in_$$GIEEE</linktorsrc></links><search><creatorcontrib>Pisa, S.</creatorcontrib><creatorcontrib>Calcagnini, G.</creatorcontrib><creatorcontrib>Cavagnaro, M.</creatorcontrib><creatorcontrib>Piuzzi, E.</creatorcontrib><creatorcontrib>Triventi, M.</creatorcontrib><creatorcontrib>Bernardi, P.</creatorcontrib><title>SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus</title><title>2006 IEEE MTT-S International Microwave Symposium Digest</title><addtitle>MWSYM</addtitle><description>Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm 3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6 degC for 6 minute MRI investigations. However, lower temperature increments are expected if blood per fusion and thermoregulation are considered</description><subject>Cardiac pacemakers</subject><subject>Catheters</subject><subject>Head</subject><subject>Humans</subject><subject>Magnetic field measurement</subject><subject>Magnetic resonance imaging</subject><subject>Pacemakers</subject><subject>Power distribution</subject><subject>Pulse measurements</subject><subject>Radio frequency</subject><subject>Temperature</subject><subject>Temperature distribution</subject><subject>Thorax</subject><issn>0149-645X</issn><isbn>9780780395411</isbn><isbn>0780395417</isbn><fulltext>true</fulltext><rsrctype>conference_proceeding</rsrctype><creationdate>2006</creationdate><recordtype>conference_proceeding</recordtype><sourceid>6IE</sourceid><sourceid>RIE</sourceid><recordid>eNotjN1KwzAYQAMqOOceQLzJC3R-SX-SXJbRucGKY5uoN478fIVqu5akE_f2ThQOHDgXh5A7BlPGQD2UL9u3csoBsilPVCyTCzJRQsKZWKUJY5dkBCxRUZakr9fkJoQPAEgly0bkfZtv6A7bHr0ejh5p0eCXHuruEGh9oGttsdWf6Omiaxz6QIvvvgvo6NDRoqTzGhsX6Np37mjP1ZxouVnSvO_17y_ckqtKNwEn_x6T53mxmy2i1dPjcpavopqJdIis5KKS0laxNRU3FrmLtTFaWYUik1I5QCcwU9YZZSqwDCxXKAWvJDdCxGNy__etEXHf-7rV_rRPgKVcqvgHp2ZVGA</recordid><startdate>200606</startdate><enddate>200606</enddate><creator>Pisa, S.</creator><creator>Calcagnini, G.</creator><creator>Cavagnaro, M.</creator><creator>Piuzzi, E.</creator><creator>Triventi, M.</creator><creator>Bernardi, P.</creator><general>IEEE</general><scope>6IE</scope><scope>6IH</scope><scope>CBEJK</scope><scope>RIE</scope><scope>RIO</scope></search><sort><creationdate>200606</creationdate><title>SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus</title><author>Pisa, S. ; Calcagnini, G. ; Cavagnaro, M. ; Piuzzi, E. ; Triventi, M. ; Bernardi, P.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-i175t-c827f88cf3cbf2bce2d3abba9c9e76889d0ed7e69cdb9bf0c10c29e872f82b773</frbrgroupid><rsrctype>conference_proceedings</rsrctype><prefilter>conference_proceedings</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Cardiac pacemakers</topic><topic>Catheters</topic><topic>Head</topic><topic>Humans</topic><topic>Magnetic field measurement</topic><topic>Magnetic resonance imaging</topic><topic>Pacemakers</topic><topic>Power distribution</topic><topic>Pulse measurements</topic><topic>Radio frequency</topic><topic>Temperature</topic><topic>Temperature distribution</topic><topic>Thorax</topic><toplevel>online_resources</toplevel><creatorcontrib>Pisa, S.</creatorcontrib><creatorcontrib>Calcagnini, G.</creatorcontrib><creatorcontrib>Cavagnaro, M.</creatorcontrib><creatorcontrib>Piuzzi, E.</creatorcontrib><creatorcontrib>Triventi, M.</creatorcontrib><creatorcontrib>Bernardi, P.</creatorcontrib><collection>IEEE Electronic Library (IEL) Conference Proceedings</collection><collection>IEEE Proceedings Order Plan (POP) 1998-present by volume</collection><collection>IEEE Xplore All Conference Proceedings</collection><collection>IEEE Electronic Library (IEL)</collection><collection>IEEE Proceedings Order Plans (POP) 1998-present</collection></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext_linktorsrc</fulltext></delivery><addata><au>Pisa, S.</au><au>Calcagnini, G.</au><au>Cavagnaro, M.</au><au>Piuzzi, E.</au><au>Triventi, M.</au><au>Bernardi, P.</au><format>book</format><genre>proceeding</genre><ristype>CONF</ristype><atitle>SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus</atitle><btitle>2006 IEEE MTT-S International Microwave Symposium Digest</btitle><stitle>MWSYM</stitle><date>2006-06</date><risdate>2006</risdate><spage>1754</spage><epage>1757</epage><pages>1754-1757</pages><issn>0149-645X</issn><isbn>9780780395411</isbn><isbn>0780395417</isbn><abstract>Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm 3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6 degC for 6 minute MRI investigations. However, lower temperature increments are expected if blood per fusion and thermoregulation are considered</abstract><pub>IEEE</pub><doi>10.1109/MWSYM.2006.249384</doi><tpages>4</tpages></addata></record>
fulltext fulltext_linktorsrc
identifier ISSN: 0149-645X
ispartof 2006 IEEE MTT-S International Microwave Symposium Digest, 2006, p.1754-1757
issn 0149-645X
language eng
recordid cdi_ieee_primary_4015289
source IEEE Electronic Library (IEL) Conference Proceedings
subjects Cardiac pacemakers
Catheters
Head
Humans
Magnetic field measurement
Magnetic resonance imaging
Pacemakers
Power distribution
Pulse measurements
Radio frequency
Temperature
Temperature distribution
Thorax
title SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T14%3A45%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-ieee_6IE&rft_val_fmt=info:ofi/fmt:kev:mtx:book&rft.genre=proceeding&rft.atitle=SAR%20Temperature%20Elevations%20in%20Pacemaker%20Holders%20Exposed%20to%20EM%20Fields%20Produced%20by%20MRI%20Apparatus&rft.btitle=2006%20IEEE%20MTT-S%20International%20Microwave%20Symposium%20Digest&rft.au=Pisa,%20S.&rft.date=2006-06&rft.spage=1754&rft.epage=1757&rft.pages=1754-1757&rft.issn=0149-645X&rft.isbn=9780780395411&rft.isbn_list=0780395417&rft_id=info:doi/10.1109/MWSYM.2006.249384&rft_dat=%3Cieee_6IE%3E4015289%3C/ieee_6IE%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_id=info:pmid/&rft_ieee_id=4015289&rfr_iscdi=true