SAR Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus
Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped...
Gespeichert in:
Hauptverfasser: | , , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm 3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6 degC for 6 minute MRI investigations. However, lower temperature increments are expected if blood per fusion and thermoregulation are considered |
---|---|