Calibration of inductive voltage dividers at extremely low frequencies
A measuring arrangement based on digital compensators is described which can be used to determine the complex division factor D_ of inductive voltage dividers (IVD) at extremely low frequencies. The error of division at 1 Hz, related to the division factor D_ at 32 Hz was evaluated as 18.7 ppm/spl p...
Gespeichert in:
Hauptverfasser: | , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A measuring arrangement based on digital compensators is described which can be used to determine the complex division factor D_ of inductive voltage dividers (IVD) at extremely low frequencies. The error of division at 1 Hz, related to the division factor D_ at 32 Hz was evaluated as 18.7 ppm/spl plusmn/3.3 ppm (1/spl sigma/) in magnitude and -4.5 m/spl deg//spl plusmn/0.42 m/spl deg/(1/spl sigma/) in phase for an IVD selected as an example.< > |
---|---|
DOI: | 10.1109/CPEM.1994.333201 |