CCE intercomparison of AC-DC transfer standards
A worldwide comparison is described covering both AC voltage and AC current measurements in the frequency range from 40 Hz to 1 MHz. Results are consistent within 10-20 p.p.m. at low voltage and frequency, but differ by up to 100 ppm at high voltage and frequency. For currents of tens of milliampere...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on instrumentation and measurement 1993-04, Vol.42 (2), p.99-108 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A worldwide comparison is described covering both AC voltage and AC current measurements in the frequency range from 40 Hz to 1 MHz. Results are consistent within 10-20 p.p.m. at low voltage and frequency, but differ by up to 100 ppm at high voltage and frequency. For currents of tens of milliamperes the results are consistent within 40 ppm.< > |
---|---|
ISSN: | 0018-9456 1557-9662 |