20-element HgI/sub 2/ energy dispersive X-ray array detector system
Recent progress in the development of HgI/sub 2/ energy dispersive X-ray detector arrays and associated miniaturized processing electronics for synchrotron radiation research applications is described. The experimental results with a 20-element array detector were obtained under realistic synchrotro...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on nuclear science 1992-10, Vol.39 (5), p.1275-1280 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Recent progress in the development of HgI/sub 2/ energy dispersive X-ray detector arrays and associated miniaturized processing electronics for synchrotron radiation research applications is described. The experimental results with a 20-element array detector were obtained under realistic synchrotron beam conditions at SSRL. An energy resolution of 250 eV at 5.9 keV was achieved. Energy resolution and throughput measurements versus input count rate and energy of the incoming radiation have been measured. Extended X-ray absorption fine structure spectra were taken from diluted samples simulating proteins with nickel.< > |
---|---|
ISSN: | 0018-9499 1558-1578 |