A readout circuit for capacitive biosensors with integrated SAR A/D conversion

This paper presents a circuit for monitoring biochemical reactions on an array of capacitive biosensors. The integrated SAR A/D conversion provides a robust solution that is configurable to meet a range of requirements. In the example given, the circuit operating at 100MHz is able to read out an arr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: van Vroonhoven, C.P.L., Rocha, D., Vellekoop, M.J., Nohammer, C.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper presents a circuit for monitoring biochemical reactions on an array of capacitive biosensors. The integrated SAR A/D conversion provides a robust solution that is configurable to meet a range of requirements. In the example given, the circuit operating at 100MHz is able to read out an array of 1000 capacitors with a value of 1pF at 12.5 kHz with 6 bit accuracy
ISSN:0271-4302
2158-1525
DOI:10.1109/ISCAS.2006.1692861