Timing performance of Hi-Rez detector for time-of-flight (TOF) PET

Positron emission tomography (PET)-based on time-of-flight (TOF) is desired to enhance the image quality by improving the noise variance in the reconstruction. This benchtop study evaluates the timing performance of the current Hi-Rez detector which is based on a 13/spl times/13 LSO array of 4 mm/sp...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 2006-06, Vol.53 (3), p.1084-1089
Hauptverfasser: Aykac, M., Bauer, F., Williams, C.W., Loope, M., Schmand, M.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Positron emission tomography (PET)-based on time-of-flight (TOF) is desired to enhance the image quality by improving the noise variance in the reconstruction. This benchtop study evaluates the timing performance of the current Hi-Rez detector which is based on a 13/spl times/13 LSO array of 4 mm/spl times/4 mm/spl times/20 mm crystals and not optimized for TOF PET applications. A Hamamatsu R9779, fast, 51-mm-diameter PMT was used for some experiments in this study as a reference PMT and its time resolution was measured to be 160 ps with a plastic scintillator. The average crystal time resolution of two Hi-Rez detectors, coupled to a 2/spl times/2 array of Hamamatsu 25-mm-diameter R8619 PMTs at +1050 V, in coincidence was measured to be 733 ps. The average crystal energy resolutions for two detectors were measured 13.1% and 13.2%. The average crystal time resolution of two Hi-Rez detectors improved from 733 ps to 642 ps when the PMT voltage was increased to +1300 V. The Hi-Rez detector was originally optimized for good energy resolution to reduce scatter and high packing fraction to obtain high sensitivity and improved spatial resolution for clinical PET.
ISSN:0018-9499
1558-1578
1558-1578
DOI:10.1109/TNS.2006.874957