Optimizing orientation of SPECT and cone beam CT detectors through quantification of cross contamination in a dual modality mammotomography system

A compact, dual modality SPECT/CT system for dedicated 3D breast imaging is in development. A major consideration in determining the relative placement of the SPECT and CT detectors is minimizing cross contamination. The transmission contamination of the emission image is investigated here using a 1...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Crotty, D.J., Bryzmialkiewicz, C.N., McKinley, R.L., Tornai, M.P.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A compact, dual modality SPECT/CT system for dedicated 3D breast imaging is in development. A major consideration in determining the relative placement of the SPECT and CT detectors is minimizing cross contamination. The transmission contamination of the emission image is investigated here using a 16/spl times/20 cm/sup 2/ quantized CZT SPECT camera located at various orientations relative to the 20/spl times/25 cm/sup 2/ CsI(TI) CT detector, with a pendant, water-filled breast phantom placed in the common field-of-view. This arrangement is repeated with and without /sup 99m/Tc emission radioactivity in the breast, with the breast laterally offset from the central ray of the X-ray cone-beam in a half-cone beam configuration. Planar transmission projections use a quasi-monochromatic (35 keV mean energy, 20% FWHM) X-ray beam, with an exposure equivalent to 1/360, 1/180, 1/90 of the total exposure of dual view, screening mammography; these fractional exposures are used due to the requirement of multiple images necessary for CT data acquisition and /spl sim/360 projection exposures are expected. Images and energy spectra acquired from the SPECT projections are used to quantify contamination from the X-ray beam. With the offset CT acquisition geometry, initial results show X-ray scatter contamination of the emission camera is limited to photons well below the 140 keV photopeak. Pulse pile-up is not a concern, with the instantaneous X-ray scatter flux. Transmission contamination in the energy windowed emission image is negligible (
ISSN:1082-3654
2577-0829
DOI:10.1109/NSSMIC.2005.1596641