Real-time estimation of oxygen concentration in micro-hemo-vessels

A real-time measurement system for non-invasive evaluation of oxygen concentration (PO/sub 2/) at the microcirculation level is developed. The system has been designed by exploiting the phenomenon of fluorescence quenching. The skin of an anaesthetized hamster, injected with porphyrin, is lighted wi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: De Francisci, M., Bucolo, M., Intaglietta, M., Arena, P., Fortuna, L.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A real-time measurement system for non-invasive evaluation of oxygen concentration (PO/sub 2/) at the microcirculation level is developed. The system has been designed by exploiting the phenomenon of fluorescence quenching. The skin of an anaesthetized hamster, injected with porphyrin, is lighted with pulses; the fluorophore reacts with the oxygen in the blood, producing a fluorescence signal, and the value of the fluorescence lifetime is related to the oxygen concentration. This microcirculation-based instrumentation consists of an electro-optical system, a control circuit and signal processing procedure. The system allows the measurement of PO/sub 2/ in the range of 0-700 (mmHg) with a standard deviation of 4 (mmHg). Several experiments have been performed in order to characterize and test this system.
DOI:10.1109/IEMBS.2004.1403650