New Calibration Technique of Contact-less Resonant Biosensor

This paper addresses the problem of calibrating magnetically-coupled passive sensors for medical applications. We developed an analytical model of such sensor including the impact of some disturbances (distance and angle between antennas). This model is valid for sensors with different dimensions of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of electronic testing 2017-06, Vol.33 (3), p.365-375
Hauptverfasser: Deluthault, Anthony, Kerzérho, Vincent, Bernard, Serge, Soulier, Fabien, Cauvet, Philippe
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This paper addresses the problem of calibrating magnetically-coupled passive sensors for medical applications. We developed an analytical model of such sensor including the impact of some disturbances (distance and angle between antennas). This model is valid for sensors with different dimensions of antennas. Thanks to this model, the ideal indicator of the disturbances for calibration can be identified. It has been applied to the case study of Intra-Ocular Pressure (IOP) sensor. A new indicator (the slope of the baseline of the Bode phase plot) has been chosen.This new indicator is reliable to significant distance and angle variations between primary and secondary.
ISSN:0923-8174
1573-0727
DOI:10.1007/s10836-017-5660-2