MUST: A silicon strip detector array for radioactive beam experiments
A new and innovative array, MUST, based on silicon strip technology and dedicated to the study of reactions induced by radioactive beams on light particles is described. The detector consists of 8 silicon strip – Si(Li) telescopes used to identify recoiling light charged particles through time of fl...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 1999-02, Vol.421 (3), p.471-491 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new and innovative array, MUST, based on silicon strip technology and dedicated to the study of reactions induced by radioactive beams on light particles is described. The detector consists of 8 silicon strip – Si(Li) telescopes used to identify recoiling light charged particles through time of flight, energy loss and energy measurements and to determine precisely their scattering angle through
X,
Y
position measurements. Each 60×60
mm
2 double sided silicon strip detector with 60 vertical and 60 horizontal strips yields an
X–
Y position resolution of 1
mm, an energy resolution of 50
keV, a time resolution of around 1
ns and a 500
keV energy threshold for protons. The backing Si(Li) detectors stop protons up to 25
MeV with a resolution of approximately 50
keV. CsI crystals read out by photo-diodes which stop protons up to 70
MeV are added to the telescopes for applications where higher energy particles need to be detected. The dedicated electronics in VXIbus standard allow us to house the 968 logic and analog channels of the array in one crate placed adjacent to the reaction chamber and fully remote controlled, including pulse visualization on oscilloscopes. A stand alone data acquisition system devoted to the MUST array has been developed. Isotope identification of light charged particles over the full energy range has been achieved, and the capability of the system to measure angular distributions of states populated in inverse kinematics reactions has been demonstrated. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/S0168-9002(98)01178-4 |