Accurate time measurements with plastic scintillation counters
A plastic scintillation counter, with a useful areaof 24×25 cm 2 and a resolution of 290 ps (fwhm) for determining the time of impact of a minimum ionizing particle, is described. The resolution may be brought down to 100 ps with smaller counters. Lower time spreads are obtained with more ionizing p...
Gespeichert in:
Veröffentlicht in: | Nuclear instruments & methods 1978-01, Vol.153 (2), p.409-412 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A plastic scintillation counter, with a useful areaof 24×25 cm
2 and a resolution of 290 ps (fwhm) for determining the time of impact of a minimum ionizing particle, is described. The resolution may be brought down to 100 ps with smaller counters. Lower time spreads are obtained with more ionizing particles. |
---|---|