Additional double-wall roof in single-wall, closed, convective incubators: Impact on body heat loss from premature infants and optimal adjustment of the incubator air temperature

Highlights • Preterm neonates lose heat by radiation to the cool surfaces of the incubator. • A double–wall roof panel in the incubator was evaluated to reduce radiant heat loss. • With an additional roof panel, radiant heat loss was lower. • However, convective and evaporative skin heat losses were...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical engineering & physics 2016-09, Vol.38 (9), p.922-928
Hauptverfasser: Delanaud, Stéphane, Decima, Pauline, Pelletier, Amandine, Libert, Jean-Pierre, Stephan-Blanchard, Erwan, Bach, Véronique, Tourneux, Pierre
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Preterm neonates lose heat by radiation to the cool surfaces of the incubator. • A double–wall roof panel in the incubator was evaluated to reduce radiant heat loss. • With an additional roof panel, radiant heat loss was lower. • However, convective and evaporative skin heat losses were higher. • This can be overcome by increasing the incubator air temperature (0.15–0.20°C).
ISSN:1350-4533
1873-4030