Near Elimination of Ventricular Pacing in SafeR Mode Compared to DDD Modes: A Randomized Study of 422 Patients
Aims: SafeR performance versus DDD/automatic mode conversion (DDD/AMC) and DDD with a 250‐ms atrioventricular (AV) delay (DDD/LD) modes was assessed toward ventricular pacing (Vp) reduction. Methods: After a 1‐month run‐in phase, recipients of dual‐chamber pacemakers without persistent AV block and...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 2012-04, Vol.35 (4), p.392-402 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aims:
SafeR performance versus DDD/automatic mode conversion (DDD/AMC) and DDD with a 250‐ms atrioventricular (AV) delay (DDD/LD) modes was assessed toward ventricular pacing (Vp) reduction.
Methods:
After a 1‐month run‐in phase, recipients of dual‐chamber pacemakers without persistent AV block and persistent atrial fibrillation (AF) were randomly assigned to SafeR, DDD/AMC, or DDD/LD in a 1:1:1 design. The main endpoint was the percentage of Vp (%Vp) at 2 months and 1 year after randomization, ascertained from device memories. Secondary endpoints include %Vp at 1 year according to pacing indication and 1‐year AF incidence based on automatic mode switch device stored episodes.
Results:
Among 422 randomized patients (73.2 ± 10.6 years, 50% men, sinus node dysfunction 47.4%, paroxysmal AV block 30.3%, bradycardia‐tachycardia syndrome 21.8%), 141 were assigned to SafeR versus 146 to DDD/AMC and 135 to DDD/LD modes. Mean %Vp at 2 months was 3.4 ± 12.6% in SafeR versus 33.6 ± 34.7% and 14.0 ± 26.0% in DDD/AMC and DDD/LD modes, respectively (P < 0.0001 for both). At 1 year, mean %Vp in SafeR was 4.5 ± 15.3% versus 37.9 ± 34.4% and 16.7 ± 28.0% in DDD/AMC and DDD/LD modes, respectively (P < 0.0001 for both). The proportion of patients in whom Vp was completely eliminated was significantly higher in SafeR (69%) versus DDD/AMC (15%) and DDD/LD (45%) modes (P < 0.0001 for both), regardless of pacing indication. The absolute risk of developing permanent AF or of remaining in AF for >30% of the time was 5.4% lower in SafeR than in the DDD pacing group (ns).
Conclusions:
In this selected patient population, SafeR markedly suppressed unnecessary Vp compared with DDD modes. PACE 2012; 35:392–402) |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/j.1540-8159.2011.03314.x |