DXA body composition corrective factors between Hologic Discovery models to conduct multicenter studies
Dual X-ray absorptiometry body composition measurements are widely used for clinical and research settings. It is well known that measurements vary across instruments, needing caution for longitudinal monitoring or multicentric studies. This study was to quantify intra- and inter-center variability...
Gespeichert in:
Veröffentlicht in: | Bone (New York, N.Y.) N.Y.), 2021-01, Vol.142, p.115683-115683, Article 115683 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Dual X-ray absorptiometry body composition measurements are widely used for clinical and research settings. It is well known that measurements vary across instruments, needing caution for longitudinal monitoring or multicentric studies. This study was to quantify intra- and inter-center variability of bone mineral content, bone mineral density, fat and lean body composition measurements between Hologic Discovery models in order to calculate the corrective factors to be applied for multicenter research projects.
A whole body phantom composed of materials representing the thickness and percentage of bone, lean and fat mass in the human physiological range was analyzed ten times in three different centers using dual energy X-ray absorptiometry scanners (Two Hologic Discovery QDR A and one QDR W). In addition, we used a morphometric vertebral phantom to monitor stability and a three steps block phantom to check accuracy.
We found a good long-term stability and accuracy for the three devices. Intra-center coefficients of variation were within the range of the manufacturer acceptable values (bone mineral density: 1.40%, bone mineral content: 1%, area: 1.50%, fat mass: 0.89%, lean mass: 0.76%, total mass: 0.12%). Whereas the inter-center coefficient of variation exceeded 8% (bone mineral density: 8.18%, bone mineral content: 3.03%, area: 8.63%: fat mass: 3,92%, lean mass: 7.89%, total mass: 2.85%).
Our study showed that the discrepancies across centers remain a major concern, particularly with regard to body composition results. Our study highlight the need of cross calibration between densitometers and proposes corrective factors evaluated from a whole body phantom to lead multicentric studies adjustment.
•Discrepancies across centers in DXA multicentric studies remain a major concern, with regard to body composition results.•We report a good long-term intra-center stability and accuracy, whereas the inter-center variability was poor.•Our study proposes corrective factors from an adapted whole body phantom to lead multicentric studies. |
---|---|