Optimization of the dwell-time for the quantitative analysis by gas chromatography-mass spectrometry
In this paper, the signal path in mass spectrometer used as a detector for gas chromatography in the selected ion monitoring mode is analyzed. The process involved an accumulation step that can be considered as a digital filter. Thus the Nyquist criterion alone is to be taken into account for the ac...
Gespeichert in:
Veröffentlicht in: | Journal of Chromatography A 1997, Vol.765 (2), p.241-245 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In this paper, the signal path in mass spectrometer used as a detector for gas chromatography in the selected ion monitoring mode is analyzed. The process involved an accumulation step that can be considered as a digital filter. Thus the Nyquist criterion alone is to be taken into account for the acquisition, by contrast to what appends with other detectors which require a higher number of acquired points to minimize the error during peak integration. This theoretical assumption was tested and accepted for three different levels of signal-to-noise ratio. |
---|---|