Prospective Randomized Crossover Study of a New Closed-loop Control System versus Pressure Support during Weaning from Mechanical Ventilation
BACKGROUND:Intellivent is a new full closed-loop controlled ventilation that automatically adjusts both ventilation and oxygenation parameters. The authors compared gas exchange and breathing pattern variability of Intellivent and pressure support ventilation (PSV). METHODS:In a prospective, randomi...
Gespeichert in:
Veröffentlicht in: | Anesthesiology (Philadelphia) 2013-09, Vol.119 (3), p.631-641 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BACKGROUND:Intellivent is a new full closed-loop controlled ventilation that automatically adjusts both ventilation and oxygenation parameters. The authors compared gas exchange and breathing pattern variability of Intellivent and pressure support ventilation (PSV).
METHODS:In a prospective, randomized, single-blind design crossover study, 14 patients were ventilated during the weaning phase, with Intellivent or PSV, for two periods of 24 h in a randomized order. Arterial blood gases were obtained after 1, 8, 16, and 24 h with each mode. Ventilatory parameters were recorded continuously in a breath-by-breath basis during the two study periods. The primary endpoint was oxygenation, estimated by the calculation of the difference between the PaO2/FIO2 ratio obtained after 24 h of ventilation and the PaO2/FIO2 ratio obtained at baseline in each mode. The variability in the ventilatory parameters was also evaluated by the coefficient of variation (SD to mean ratio).
RESULTS:There were no adverse events or safety issues requiring premature interruption of both modes. The PaO2/FIO2 (mean ± SD) ratio improved significantly from 245 ± 75 at baseline to 294 ± 123 (P = 0.03) after 24 h of Intellivent. The coefficient of variation of inspiratory pressure and positive end-expiratory pressure (median [interquartile range]) were significantly higher with Intellivent, 16 [11–21] and 15 [7–23]%, compared with 6 [5–7] and 7 [5–10]% in PSV. Inspiratory pressure, positive end-expiratory pressure, and FIO2 changes were adjusted significantly more often with Intellivent compared with PSV.
CONCLUSIONS:Compared with PSV, Intellivent during a 24-h period improved the PaO2/FIO2 ratio in parallel with more variability in the ventilatory support and more changes in ventilation settings. |
---|---|