Measurement and evaluation of galvanic corrosion between titanium/Ti6Al4V implants and dental alloys by electrochemical techniques and auger spectrometry
The purpose of this study was to investigate, in different experimental conditions, the galvanic corrosion phenomena which can exist between a dental suprastructure and a dental implant. The electrochemical behavior of 7 alloy superstructures with titanium and titanium alloy (Ti6Al4V) implants was i...
Gespeichert in:
Veröffentlicht in: | Biomaterials 1999-05, Vol.20 (10), p.933-941 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to investigate, in different experimental conditions, the galvanic corrosion phenomena which can exist between a dental suprastructure and a dental implant. The electrochemical behavior of 7 alloy superstructures with titanium and titanium alloy (Ti6Al4V) implants was investigated by electrochemical means in Fusayama–Meyer de-aerated saliva and Carter–Brugirard (AFNOR) non de-aerated saliva. Different techniques were used to obtain the value of the galvanic coupling current and potential for each couple. All showed very low corrosion rates, ranging from 10
-6 to 10
-8
A. Surface analysis confirmed these results. |
---|---|
ISSN: | 0142-9612 1878-5905 |
DOI: | 10.1016/S0142-9612(98)00248-8 |