Accuracy of the robot‐assisted procedure in deep brain stimulation
Introduction The use of a robot‐assisted technology becomes very competitive. The aim of this work was to define the accuracy of robotic assistance in deep brain stimulation surgery and to compare results with that in the literature. Methods We retrospectively reviewed the accuracy of lead implantat...
Gespeichert in:
Veröffentlicht in: | The international journal of medical robotics + computer assisted surgery 2019-12, Vol.15 (6), p.e2032-n/a |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
The use of a robot‐assisted technology becomes very competitive. The aim of this work was to define the accuracy of robotic assistance in deep brain stimulation surgery and to compare results with that in the literature.
Methods
We retrospectively reviewed the accuracy of lead implantation in 24 consecutive patients who had robot‐assisted (ROSA, Zimmer‐Biomet) surgery for the treatment of movement disorders. Intended stereotactic coordinates (x, y, z) of contact 0 (the most distal contact at the tip of the electrode) of each definitive lead were compared with actual coordinates obtained by a postoperative CT scan. For each lead, the euclidian 3D distance between the actual and intended location of contact 0 was calculated.
Results
The euclidian 3D distances between the intended and actual location of the contact 0 were 0.81 mm on the right side and 1.12 mm on the left side.
Discussion
Robot‐assisted technology for stereotactic surgery is safe and accurate. The association with a 3D flat‐panel CT scan is an optimized procedure for deep intracranial electrode implantation. |
---|---|
ISSN: | 1478-5951 1478-596X |
DOI: | 10.1002/rcs.2032 |