Interest of intra-operative 3D imaging in spine surgery: a prospective randomized study

Purpose We report a single-center, prospective, randomized study for pedicle screw insertion in opened and percutaneous spine surgeries, using a computer-assisted surgery (CAS) technique with three-dimensional (3D) intra-operative images intensifier (without planification on pre-operative CT scan) v...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European spine journal 2016-06, Vol.25 (6), p.1738-1744
Hauptverfasser: Ruatti, Sébastien, Dubois, C., Chipon, E., Kerschbaumer, G., Milaire, M., Moreau-Gaudry, A., Tonetti, J., Merloz, Ph
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose We report a single-center, prospective, randomized study for pedicle screw insertion in opened and percutaneous spine surgeries, using a computer-assisted surgery (CAS) technique with three-dimensional (3D) intra-operative images intensifier (without planification on pre-operative CT scan) vs conventional surgical procedure. Material and method We included 143 patients: Group C (conventional, 72 patients) and Group N (3D Fluoronavigation, 71 patients). We measured the pedicle screw running time, and surgeon’s radiation exposure. All pedicle runs were assessed according to Heary by two independent radiologists on a post-operative CT scan. Results 3D Fluoronavigation appeared less accurate in percutaneous procedures (24 % of misplaced pedicle screws vs 5 % in Group C) ( p  = 0.007), but more accurate in opened surgeries (5 % of misplaced pedicle screws vs 17 % in Group C) ( p  = 0.025). For one vertebra, the average surgical running time reached 8 min in Group C vs 21 min in Group N for percutaneous surgeries ( p  = 3.42 × 10 −9 ), 7.33 min in Group C vs 16.33 min in Group N ( p  = 2.88 × 10 −7 ) for opened surgeries. The 3D navigation device delivered less radiation in percutaneous procedures [0.6 vs 1.62 mSv in Group C ( p  = 2.45 × 10 −9 )]. For opened surgeries, it was twice higher in Group N with 0.21 vs 0.1 mSv in Group C ( p  = 0.022). Conclusion The rate of misplaced pedicle screws with conventional techniques was nearly the same as most papers and a little bit higher with CAS. Surgical running time and radiation exposure were consistent with many studies. Our work hypothesis is partially confirmed, depending on the type of surgery (opened or closed procedure).
ISSN:0940-6719
1432-0932
DOI:10.1007/s00586-015-4141-5