Determination of Cu and Mn in seawater by graphite furnace atomic absorption spectrometry with the use of hydrofluoric acid as a chemical modifier

The addition of hydrogen fluoride to a seawater sample in a graphite tube atomizer makes possible the removal of the major matrix components in an optimized pre-heating step, thus a near interference free determination of copper and manganese trace metals is achieved. The optimization studies were s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Spectrochimica acta. Part B: Atomic spectroscopy 2002-05, Vol.57 (5), p.939-950
1. Verfasser: Cabon, J.Y.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The addition of hydrogen fluoride to a seawater sample in a graphite tube atomizer makes possible the removal of the major matrix components in an optimized pre-heating step, thus a near interference free determination of copper and manganese trace metals is achieved. The optimization studies were supported by a priori calculation of limit of detection (LOD) data. For optimized conditions, the experimentally determined LOD values (3σ, n=20) were 0.11 and 0.03 μg l −1 for Cu and Mn, respectively, for 20 μl applied seawater volume. With the use of calibration with standard addition, determined concentrations of Cu and Mn in CASS-2 seawater and in SLEW-1 estuarine water were in good agreement with certified values at the 95% confidence level ( n=10).
ISSN:0584-8547
1873-3565
DOI:10.1016/S0584-8547(02)00025-3